Temperature (English) Operating-40°F to +122°F
Temperature (Metric) Operating-40°C to +50°C
CalibrationDirect software support for single-point and multi-point audit/calibrations of the volumetric flow
Certifications/ComplianceCE: EN550011 Group 1, Class B (Emissions); EN55082-1 (Immunity); EN61010-1 (Safety)ETL: UL and CSA Equivalent approval
Data Points86 days of Interval data (stored every 5 minutes), 32 filter data records, and input data (30 minute default)
Depth (English)15.8 in.
Depth (Metric)40.2 cm
DescriptionAir Sampler
Electrical Requirements3 A (at 120 VAC); 1.5 A (at 240 VAC)
Filter Holders and MediaAccommodates 47 mm filter in the single-flow and dual-flow configurations (for USEPA PM-2.5 and PM-Coarse sampling); PTFE 2 μm pore size membrane filter (regulation required) (for USEPA PM-10 reference sampling: Pallflex TX-40, quartz filter and PTFE
Height (English)26.3 in. (Overall)/31 in. (Top Cover)/27.1 in. (Stand)
Height (Metric)67.3 cm (Overall)/78.8 cm (Top Cover)/68.8 cm (Stand)
OutputsKeypad/display for data retrieval and user programming, RS232/RS485 interfaceWind vane/anemometer connection with 24Vdc power output and 2 0-5 Vdc inputs for wind speed and directionDual USB ports for easy data download to universal flash drivesOptional kit available to configure the sampler for digital and analog I/O to enable connection to other devices
Sample Flow Control and ReportingActivol flow control system using mass flow controller. Controls primary (PM-2.5) sample flow at 15 L/min and secondary (PM-Coarse) sample flow at 1.67 L/min. The sampler displays the current volumetric flow rate (L/min.).
Weight (English)101 lb.
Weight (Metric)46 kg
Width (English)25.2 in. (Overall)/35.3 in. (Inlet Connectors)/30.8 in. (Stand)/42 in. (Footprint)
Width (Metric)64; 89.5; 78.2; 106.7 cm (Overall/Inlet Connector/Stand/Footprint)