Specifications
Flow Rate1 to 3.5 L/min.
Accuracy±5% of reading ± precision (traceable to SAE Fine Test Dust)
Aerodynamic Particle Cut-Point Range1.0 to 10 μm
Certifications/ComplianceSuitable for NIOSH Methods 0500 and 0600. CE marked. UKCA
Concentration Display Averaging Time1 to 60 seconds (user selectable)
Concentration Display Updating Interval1 second
Concentration Measurement Range0.001 to 400mg/m3 range (auto ranging)1
Data Points>500,000
Depth (English)3.3 in.
Depth (Metric)84 mm
DescriptionpersonalDataRAM pDR-1500 Aerosol Monitor
Height (English)7.1 in.
Height (Metric)181 mm
Particle Size Range0.1 to 10µm
Precision±2% of reading or ±0.005 mg/m3, whichever is larger, for 1 second (2-sigma)2 averaging time, ±0.5% of reading or ±0.0015 mg/m3, whichever is larger, for 10 second averaging time, ±0.2% of reading or ±0.0005 mg/m3, whichever is larger, for 60 second averaging time
Resolution0.1% of reading or 0.001 mg/m3, whichever is larger
Scattering Coefficient Range1.5 x 10-6 to 0.6 m-1 (approx.) @ lambda= 880nm (not displayed)
Weight (English)41 oz.
Weight (Metric)1.2 kg
Width (English)5.6 in.
Width (Metric)143 mm
Unit SizeEach