Search Thermo Fisher Scientific
The TaqMan RNase P Instrument Verification Plate is used to verify the performance of the Applied Biosystems 7900HT 96-Well Real-Time PCR System and the ABI PRISM 7000 and 7700 Sequence Detection Systems. The TaqMan RNase P Instrument Verification Plate is a 96-well PCR plate preloaded with the necessary reagents to detect and quantitate genomic copies of the human RNase P gene, a single-copy gene encoding the RNA moiety of the RNase P enzyme. Each well contains preloaded reaction mix (1X TaqMan Universal PCR Master Mix, RNase P primers, and FAM™ dye–labeled probe) and template.
The TaqMan RNase P Instrument Verification Plate satisfies the same requirements as the Applied Biosystems 7900HT 96-Well Real-Time PCR System, and the ABI PRISM 7000 and 7700 Sequence Detection Systems installations. To pass installation, the instruments must demonstrate the ability to distinguish between 5,000 and 10,000 genomic equivalents with a 99.7% confidence level for a subsequent sample run in a single well. To ensure accurate results, Applied Biosystems recommends that the performance of your instrument be verified every six months.